气泡内水汽平衡研究与气泡法饱和盐湿度发生器的原理
The Equilibrium of the Water Vapor in an Air Bubble and the Principle of the Humidity Generator with Saturated Aqueous Salt Solution by Means of the Air Bubbles
-
摘要: 气泡具有密封性好、其内部水汽平均扩散距离小的特点, 因此, 其内部很容易达到水汽平衡, 利用这个特点的气泡法饱和盐湿度发生器, 与通常饱和盐湿度发生器比较具有效率高的突出优点。该文重点定量分析了气泡内水汽平衡条件; 定性分析了在发生湿度过程中其他重要物理效应。利用General Eastern公司M4型露点仪作为标准器, 实测试验装置输出湿度与试验装置标称湿度之间差值的绝对值不大于1.0%。Abstract: The constant humidity property of some saturated saline solution can be used to design humidity generators. Current humidity generators using saturated saline solution usually require complex equipments and a long time to reach equilibrium, and are prone to environmental influences. On the other hand, air bubbles are characterized by good air-tightness and short inside mean diffusion distance, which is defined as the volume above the solution divided by the surface area of the solution in a closed container. Therefore, it is easier to achieve vapor equilibrium within an air bubble, which can then be collected pneumatically and used to calibrate humidity sensors.The conditions under which water vapor reaches equilibrium within air bubbles are quantitatively analyzed in this study. An air bubble can be modeled as a closed cylinder container with solution at the bottom. Assuming that the gradient of vapor density is distributed linearly along the axis of the cylinder, the time τ required to reach vapor equilibrium within the air bubble can be deduced from the relationship that the vapor mass diffused from the solution surface into the container within a unit time equals the increased vapor mass within the container. From the forces exerted on the air bubble during its ascent, the time t for the air bubble to remain in the solution can be calculated. In order to reach vapor equilibrium within the air bubble, τ must be smaller than t (τ < t). It follows that the diameter of the air bubble r and the depth of the solution h should satisfy the inequality below: in which ρ is the solution density, g is the gravity acceleration, D is the vapor diffusion constant, η is the viscosity of the solution, U0 is the normal humidity at the surface of the saturated solution, U1 is the initial humidity in the air bubble, and U2 is the humidity of the air released from the air bubble when it bursts.To test the new humidity generator described above, a test set is calibrated against a M4 hygrometer (General Eastern Instrument) with the dew point temperature uncertainty of 0.2 ℃. The absolute values of the differences between readings from the test humidity generator and those from the reference hygrometer are no bigger than 1%. The response time is typically 10 minutes under common application conditions.The sources of error for the humidity generator using the air bubble method are also discussed. The analysis and the test show that a humidity generator based on air bubbles from saturated saline solution is simple, fast, accurate, and less affected by environmental factors, and is a real utility humidity reference, in particular can be used in onsite calibration of humidity sensors at automatic weather stations.
-
Key words:
- humidity generator;
- saturated saline solution;
- air bubble
-
表 1 几种盐的饱和溶液在常温下的相对湿度值
表 2 气泡法饱和盐湿度发生器试验装置标称湿度与实测值
-
[1] 李英干, 范金鹏.湿度测量.北京:气象出版社, 1990: 350-361; 503-505. [2] 王金钊.恒湿盐湿度发生器研制.应用气象学报, 1995, 6 (4): 427-435. http://qikan.camscma.cn/jams/ch/reader/view_abstract.aspx?file_no=19950466&flag=1 [3] 徐祝龄.气象学.北京:气象出版社, 1994: 384-387.